1997 Revised Guidelines for Performing CD4+ T-Cell Determinations in Persons Infected with Human Immunodeficiency Virus (HIV)Recommendations
January 10, 1997
I. Laboratory Safety
B. Establish the following safety practices (38-44):
B. Collect blood specimens by venipuncture (54) into evacuated tubes containing an appropriate anticoagulant, completely expending the vacuum in the tubes.
C. Draw the appropriate number of tubes:
D. Label all specimens with the date, time of collection, and a unique patient identifier.
A. Maintain and transport specimens at room temperature (64-72 F [18-22 C]) (52, 55-57). Avoid extremes in temperature so that specimens do not freeze or become too hot. Temperatures >99 F (37 C) may cause cellular destruction and affect both hematology and flow cytometry measurements (52). In hot weather, packing the specimen in an insulated container and placing this container inside another containing an ice pack and absorbent material may be necessary. This method helps retain the specimen at ambient temperature. The effect of cool temperatures (i.e., 39 F [4 C]) on immunophenotyping results is not clear (52, 57). B. Transport specimens to the immunophenotyping laboratory as soon as possible. C. For transport to locations outside the collection facility but within the state, follow state or local guidelines. One method for packaging such specimens is to place the tube containing the specimen in a leak-proof container (e.g., a sealed plastic bag) and to pack this container inside a cardboard canister containing sufficient material to absorb all the blood should the tube break or leak. Cap the canister tightly. Fasten the request slip securely to the outside of this canister with a rubber band. For mailing, this canister should be placed inside another canister containing the mailing label. D. For interstate shipment, follow federal guidelines* for transporting diagnostic specimens. Note: Use overnight carriers with an established record of consistent overnight delivery to ensure arrival the following day. Check with these carriers for their specific packaging requirements. E. Obtain specific protocols and arrange appropriate times of collection and transport from the facility collecting the specimen. A. Inspect the tube and its contents immediately upon arrival. B. Take corrective actions if the following occur: 1. If the specimen is hot or cold to the touch but not obviously hemolyzed or frozen, process it but note the temperature condition on the work-sheet and report form. Do not rapidly warm or chill specimens to bring them to room temperature because this may adversely affect the immunophenotyping results (52). Abnormalities in light-scattering patterns will reveal a compromised specimen. 2. If blood is hemolyzed or frozen, reject the specimen and request another. 3. If clots are visible, reject the specimen and request another. 4. If the specimen is >48 hours old (from the time of draw), reject it and request another.
1. Perform the hematologic tests within the time frame specified by the manufacturer of the specific hematology instrument used (time from blood specimen draw to hematologic test). (See Note under II.A.1.b.) 2. Perform an automated WBC count and differential, counting 10,000- 30,000 cells (58). If the specimen is rejected or "flagged" by the instrument, a manual differential of at least 400 cells can be performed. If the flag is not on the lymphocyte population and the lymphocyte differential is reported by the instrument, the automated lymphocyte differential should be used. 3. If absolute counts are determined by using a single-platform method, hematology results are not needed for this determination. B. Immunophenotyping 1. For optimal results, perform the test within 30 hours, but no later than 48 hours, after drawing the blood specimen (59, 60). 2. When centrifuging, maintain centrifugation forces of no greater than 400 g for 3-5 minutes for wash steps. 3. Vortex sample tubes to mix the blood and reagents and break up cell aggregates. Vortex samples immediately before analysis to optimally disperse cells. 4. Include a source of protein (e.g., fetal bovine serum or bovine serum albumin) in the wash buffer to reduce cell clumps and non-specific fluorescence. 5. Incubate all tubes in the dark during the immunophenotyping procedure. 6. Before analysis on the flow cytometer, be sure all samples have been adequately fixed. Although some of the commercial lysing/fixing reagents can inactivate cell-associated HIV, all tubes should be fixed after staining and lysing with 1%-2% buffered paraformaldehyde or formaldehyde. Note: The characteristics of paraformaldehyde and formaldehyde may vary between lots. They may also lose their effectiveness over time. Therefore, these fixatives should be made fresh weekly from electron-microscopy- grade aqueous stock. 7. Immediately after processing the specimens, store all stained samples in the dark and at refrigerator temperatures (39-50 F [4-10 C]) until flow cytometric analysis. These specimens should be stored for no longer than 24 hours unless the laboratory can demonstrate that scatter and fluorescence patterns do not change for specimens stored longer. 8. If absolute counts are determined on the flow cytometer, follow the manufacturer's recommended protocols.
1. CD4 T-cells must be identified as being positive for both CD3 and CD4. 2. CD8 T-cells must be identified as being positive for both CD3 and CD8. B. Two-color monoclonal antibody panels
a. CD3 Monoclonal antibody in tubes 3-6 serves as a control for tube-to-tube variability and is also used to determine T-cell populations. Note: All CD3 values in this six-tube panel should be within 3% of each other. If the CD3 value of a tube is >3% of any of the others, that tube should be repeated (i.e., new aliquot of blood labeled, lysed, and fixed). b. Monoclonal antibodies that label T-cells, B-cells, and NK-cells are used to account for all lymphocytes in the specimen (61).
2. An abbreviated two-color panel should only be used for testing specimens from patients for whom CD4+ T-cell levels are being requested as part of sequential follow-up, and then only after consulting with the requesting clinician. Because some of the internal controls are no longer included, when using an abbreviated panel, the immunophenotyping results should be reviewed carefully to ensure that CD3+ T-cell levels are similar to those determined previously with the full recommended panel. When discrepancies occur, the specimens must be reprocessed by using the full recommended two-color monoclonal antibody panel. C. Three-color monoclonal antibody panels 1. Three-color monoclonal antibody panels should fulfill the following basic requirements: enumerate CD4+ and CD8+ T-cells, validate the lymphocyte gate used, and provide some assessment of tube-to-tube variability. 2. For determining T-cell subset percentages, the third color should be used to identify lymphocytes by following one of two procedures (Table 2): a. Use CD45 as the third color to identify lymphocytes as those cells that are bright CD45+ but have low side scattering properties. In this case, the panel would consist of the following monoclonal antibodies: CD3/CD4/CD45; CD3/CD8/CD45; and CD3/CD19/CD45 (Table 2, Panel A). b. Use lineage markers (T-cell, B-cell, and NK-cell) to identify lymphocytes (63). The panel would consist of the following monoclonal antibodies: CD3/CD19/CD16 and/or CD56; CD3/CD4/CD8; and an isotype control (Table 2, Panel B). Note: Software on the flow cytometer must be capable of using the information obtained from these monoclonal antibody combinations to correctly identify lymphocytes and to extrapolate that information to determine the percentage of lymphocytes that are CD4+ and CD8+ T-cells (63). Note: A single tube containing CD3, CD4, and CD8 monoclonal antibodies is not appropriate for determining the percentage of lymphocytes that are CD4+ or CD8+ T-cells because there is no method to validate the lymphocyte gate in this tube without the addition of another tube for that purpose. Lymphocyte gate purity and recovery cannot be determined. Internal quality control measures may be obtained by adding another tube containing CD3 (e.g., CD3, CD19, and CD16 and/or CD56). D. Four-color monoclonal antibody panels 1. Addition of CD45 to a single tube containing CD3, CD4, and CD8 allows the identification of lymphocytes based on CD45 and side scatter and the enumeration of CD4+ and CD8+ T-lymphocytes. 2. A four-color monoclonal antibody panel must consist of at least two tubes, each with the same lineage marker. A second tube containing CD45, CD3, CD19, and CD16 and/or CD56 is recommended. A. Negative (isotype) reagent control 1. Use this control to determine nonspecific binding of the mouse monoclonal antibody to the cells and to set markers for distinguishing fluorescence-negative and fluorescence-positive cell populations. 2. Use a monoclonal antibody with no specificity for human blood cells but of the same isotype(s) as the test reagents. Note: In many cases, the isotype control may not be optimal for controlling nonspecific fluorescence because of differences in F/P ratio, antibody concentration between the isotype control and the test reagents, and other characteristics of the immunoglobulin in the isotype control. Additionally, isotype control reagents from one manufacturer are not appropriate for use with test reagents from another manufacturer. 3. The isotype control is not needed for use with CD45 because CD45 is used to identify leukocyte populations based on fluorescence intensity. 4. For monoclonal antibody panels containing antibodies to CD3, CD4, and CD8, the isotype control may not be needed because labeling with these antibodies results in fluorescence patterns in which the unlabeled cells are clearly separated from the labeled cells. In these instances, the negative cells in the histogram are the appropriate isotype control. 5. The isotype control must be used when a monoclonal antibody panel contains monoclonal antibodies that label populations that do not have a distinct negative population (e.g., some CD16 or CD56 monoclonal antibodies). B. Positive methodologic control 1. The methodologic control is used to determine whether procedures for preparing and processing the specimens are optimal. This control is prepared each time specimens from patients are prepared. 2. Use either a whole-blood specimen from a control donor or commercial materials validated for this purpose. Ideally, this control will match the population of patients tested in the laboratory. (See Section XII.D.) 3. If the methodologic control falls outside established normal ranges, determine the reason. Note: The purpose of the methodologic control is to detect problems in preparing and processing the specimens. Biologic factors that cause only the whole-blood methodologic control to fall outside normal ranges do not invalidate the results from other specimens processed at the same time. Poor lysis or poor labeling in all specimens, including the methodologic control, invalidates results. C. Positive control for testing reagents 1. Use this control to test the labeling efficiency of new lots of reagents or when the labeling efficiency of the current lot is questioned. Prepare this control only when needed (i.e., when reagents are in question) in parallel with lots of reagents of known acceptable performance. Note: New reagents must demonstrate similar results to those of known acceptable performance. 2. Use a whole-blood specimen or other human lymphocyte preparation (e.g., cryopreserved or commercially obtained lyophilized lymphocytes). A. Align optics daily. This ensures that the brightest and tightest peaks are produced in all parameters. Note: Some clinical flow cytometers can be aligned by laboratory personnel whereas others can be aligned only by qualified service personnel. 1. Align the flow cytometer by using stable calibration material (e.g., microbeads labeled with fluorochromes) that has measurable forward scatter, side scatter, and fluorescence peaks. 2. Align the calibration particles optimally in the path of the laser beam and in relation to the collection lens so the brightest and tightest peaks are obtained. 3. Align stream-in-air flow cytometers daily (at a minimum) and stream-in-cuvette flow cytometers (most clinical flow cytometers are this type) as recommended by the manufacturer. B. Standardize daily. This ensures that the flow cytometer is performing optimally each day and that its performance is the same from day to day. 1. Select machine settings that are optimal for fluorochrome-labeled, whole-blood specimens. 2. Use microbeads or other stable standardization material to place the scatter and fluorescence peaks in the same scatter and fluorescence channels each day. Adjust the flow cytometer as needed. 3. Maintain records of all daily standardizations. Monitor these to identify any changes in flow cytometer performance. 4. Retain machine standardization settings for the remaining quality control procedures (sensitivity and color compensation) and for reading the specimens. C. Determine fluorescence resolution daily. The flow cytometer must differentiate between the dim peak and autofluorescence in each fluorescence channel. 1. Evaluate standardization/calibration material or cells that have low-level fluorescence that can be separated from autofluorescence (e.g., microbeads with low-level and negative fluorescence or CD56-labeled lymphocyte preparation). 2. Establish a minimal acceptable distance between peaks, monitor this difference, and correct any daily deviations. D. Compensate for spectral overlap daily. This step corrects the spectral overlap of one fluorochrome into the fluorescence spectrum of another. 1. Use either microbead or cellular compensation material containing three populations for two-color immunofluorescence (no fluorescence, PE fluorescence only, and FITC fluorescence only), four populations for three-color immunofluorescence (the three above plus a population that is positive for only the third color), or five populations for four-color (the four above plus a population that is positive for only the fourth color). 2. Analyze this material and adjust the electronic compensation circuits on the flow cytometer to place the fluorescent populations in their respective fluorescence quadrants with no overlap into the double-positive quadrant (Figure 1). If three fluorochromes are used, compensation must be carried out in an appropriate sequence: FITC, PE, and the third color, respectively (64). For four-color monoclonal antibody panels, follow the flow cytometer manufacturer's instructions for four fluorochromes. Avoid overcompensation. 3. If standardization or calibration particles (microbeads) have been used to set compensation, confirm proper calibration by using lymphocytes labeled with FITC- and PE-labeled monoclonal antibodies (and a third-color- or fourth-color-labeled monoclonal antibody for three-color or four-color panels) that recognize separate cell populations but do not overlap. These populations should have the brightest expected signals. Note: If a dimmer-than-expected signal is used to set compensation, suboptimal compensation for the brightest signal can result. 4. Reset compensation when photomultiplier tube voltages or optical filters are changed. E. Repeat all four instrument quality control procedures whenever instrument problems occur or if the instrument is serviced during the day. F. Maintain instrument quality-control logs, and monitor them continually for changes in any of the parameters. In the logs, record instrument settings, peak channels, and coefficient of variation (CV) values for optical alignment, standardization, fluorescence resolution, and spectral compensation. Reestablish fluorescence levels for each quality-control procedure when lot numbers of beads are changed. IX. Sample AnalysesA. For the two-color immunophenotyping panel using a light-scatter gate, analyze the sample tubes of each patient's specimen in the following order: 1) The tube containing CD45 and CD14 (gating reagent): read this tube first so that gates can be set around the lymphocyte cluster; 2) Isotype control: set cursors for differentiating positive and negative populations so that £2% of the cells are positive; and 3) Remaining tubes in the panel. 1. Count at least 2,500 gated lymphocytes in each sample. This number ensures with 95% confidence that the result is £2% standard deviation (SD) of the "true'' value (binomial sampling). Note: This model assumes that variability determined from preparing and analyzing replicates is £2% SD. Each laboratory must determine the level of variability by preparing and analyzing at least eight replicates of the last four tubes in the recommended panel. Measure variability when first validating the methodology used and again when methodologic changes are made. 2. Examine light-scattering patterns on each sample tube. Determine whether lysis or sample preparation, which can affect light scattering, is the same in each sample tube of a patient's specimen. Deviation in a particular tube usually indicates sample preparation error, and the tube should be repeated (i.e., a new aliquot of blood should be stained and lysed).FIGURE 1. Determination of appropriate compensation for specral overlap B. For three- or four-color monoclonal antibody panels using a CD45/side scatter gate, determine the lymphocyte population based on bright CD45 fluorescence and low side scattering properties. Draw a gate on this population and analyze the cell populations using this gate (65). X. Data AnalysisA. Light-scatter gate (for two-color panels). 1. Reading from the sample tube containing CD45 and CD14, draw lymphocyte gates using forward and side light-scattering patterns and fluorescence staining.
a. When using CD45 and CD14 and light-scattering patterns for drawing lymphocyte gates, define populations on the following basis:
b. Using the above characteristics, draw a light-scattering gate around the lymphocyte population (66). Note: Other methods for drawing a lymphocyte gate must accurately identify lymphocytes and account for non-lymphocyte contamination of the gate. 2. Verify the lymphocyte gate by determining the recovery of lymphocytes within the gate and the lymphocyte purity of the gate.
a. Definitions
b. Optimally, the lymphocyte recovery should be 95%. c. Optimally, the lymphocyte purity of the gate should be 90%. d. Optimal gates include as many lymphocytes and as few contaminants as possible. e. Lymphocyte recovery within the gate using CD45 and CD14 can be determined by two different methods: light-scatter gating and fluorescence gating (Figures 2 and 3). Note: The number of lymphocytes identified will be the same whether determined by light-scatter gating or by fluorescence gating.
f. The lymphocyte purity of the gate is determined from the CD45 and CD14 tube by calculating the percentage of cells in the light-scattering gate that are bright CD45-positive and negative for CD14. g. If the recommended recovery and purity of lymphocytes within the gate cannot be achieved, redraw the gate. If minimum levels still cannot be obtained, reprocess the specimen. If this fails, request another specimen. B. CD45 gating (for three- and four-color monoclonal panels) 1. Identify lymphocytes as cells brightly labeled with CD45 and having low side scattering properties. 2. Establish criteria for cluster identification based on a clear definition of lymphocytes that does not include basophils (less bright CD45, low side scatter) or monocytes (less bright CD45, moderate side scatter). Note: Care must be taken to include all lymphocytes. B-cells may have slightly less CD45 fluorescence than the T-cells (the major cluster of lymphocytes). NK-cells have bright CD45 fluorescence but have slightly more side scattering properties than the majority of the lymphocytes.FIGURE 2. Light-scatter gating technique for determining lymphocyte recovery and purity 3. CD45/side scatter gates for lymphocytes are assumed to contain >95% lymphocytes, and no further corrections need be made to the percentage subset results ( 65 ). 4. Lymphocyte recovery cannot be determined without using a panel of monoclonal antibodies that identify T-, B-, and NK-cells. Note: Validation of a CD45/side scatter gate is recommended when beginning to use CD45/ side scatter gates to help determine the CD45 and side scatter characteristics of T-, B-, and NK-cells and to ensure their inclusion in the gate. C. Set cursors using the isotype control so that <2% of cells are positive. Note: If an isotype control is not used, set cursors based on the tube containing CD3 and CD4 so that the negative and positive cells in the histogram are clearly separated. These cursors may be used for the remaining tubes. If CD16 and/or CD56 are included in a monoclonal antibody panel, an isotype control may be needed to help identify negative cells. FIGURE 3. Flourescence gating technique for determining lymphocyte recovery and purity D. Analyze the remaining samples with the cursors set. Note: In some instances, the isotype-set cursors will not accurately separate positive and negative staining for another sample tube from the same specimen. In such cases, the cursors can be moved on that sample to more accurately separate these populations. The cursors should not be moved when fluorescence distributions are continuous with no clear demarcation between positively and negatively labeled cells. E. Analyze each patient or control specimen with lymphocyte gates and cursors for positivity set for that particular patient or control. F. When spectral compensation of a particular specimen appears to be inappropriate because FITC-labeled cells have been dragged into the PE-positive quadrant or vice-versa (when compensation on all other specimens is appropriate) (67), repeat the sample preparation, prewashing the specimen with phosphate-buffered saline (PBS) (pH 7.2) to remove plasma before monoclonal antibodies are added. G. Include the following analytic reliability checks, when available: 1. Optimally, at least 95% lymphocyte recovery (proportion of lymphocytes within the lymphocyte gate) should be achieved. Minimally, at least 90% lymphocyte recovery should be achieved. Note: These determinations can only be made when using either CD14 and CD45 to validate the gate or when using T, B, and NK reagents to validate a gate. 2. Optimally, 90% lymphocyte purity should be observed within the lymphocyte gate. Minimally, 85% purity should be observed within the gate.
3. Optimally, the sum of the percentage of CD3+CD4+ and CD3+CD8+
cells
should equal the total percentage of CD3+ cells within ±5%, with
a maximum
variability of £10%. Note: In specimens containing a
considerable
number of T 4. Optimally, the sum of the percentage of CD3+ (T-cells), CD19+ (B-cells), and CD3-(CD16 and/or CD56)+ (NK-cells) should equal the purity of lymphocytes in the gate ±5% (61), with a maximum variability of £10%. If the data are corrected for lymphocyte purity (see XII.B.), the sum should ideally equal 95%-105% (or at a minimum 90%-110%). A. If possible, store list-mode data on all specimens analyzed. This allows for reanalysis of the raw data, including redrawing of gates. At a minimum, retain hard copies of the lymphocyte gate and correlated dual histogram data of the fluorescence of each sample. B. Retain all primary files, worksheets, and report forms for 2 years or as required by state or local regulation, whichever is longer. Data can be stored electronically. Disposal after the retention period is at the discretion of the laboratory director. A. Report all data in terms of CD designation, with a short description of what that designation means. Note: CD4+ T-cells are T-helper cells. The correct cells to report for this value are those that are positive for both CD3 and CD4. Similarly, CD8+ T-cells are T-suppressor/cytotoxic cells and are positive for both CD3 and CD8. Do not include other cell types (non-T-cells) in CD4 and CD8 T-cell determinations. B. If using light-scatter gates, report data as a percentage of the total lymphocytes and correct for the lymphocyte purity of the gate. For example, if the lymphocyte purity is 94% and the CD3 value is 70%, correct the CD3 value by dividing 0.70 by 0.94 and then multiply the result by 100 to result in a T-lymphocyte value of 74%. C. Report absolute lymphocyte subset values when an automated complete blood cell (CBC) count (WBC and differential) has been performed from blood drawn at the same time as that for immunophenotyping. 1. Calculate the absolute values by multiplying the lymphocyte subset percentage (from flow cytometry data) by the absolute number of lymphocytes (from WBC and differential). Note: The hematology laboratory providing the CBC (WBC and differential) must perform satisfactorily in a hematology proficiency testing program approved by the Health Care Finance Administration (HCFA) as meeting the requirements of the Clinical Laboratory Improvement Amendments of 1988 (CLIA '88).** 2. Report both percentages and absolute counts when these are available. Note: If absolute counts are determined directly on the flow cytometer, report these results. D. Report data from all relevant monoclonal antibody combinations with corresponding reference limits of expected normal values (e.g., CD4+ T-cell percentage and absolute number of CD4+ T-cells). Reference limits for immunophenotyping test results must be determined for each laboratory (29). Separate reference ranges must be established for adults and children, and the appropriate ranges must be used for patient specimens. XIII.Quality Assurance
1. Methods for collecting, handling, transporting, identifying, processing, and storing specimens. 2. Information provided on test request and results report forms. 3. Instrument performance, quality-control protocols, and maintenance. 4. Reagent quality-control protocols. 5. Process for reviewing and reporting results. 6. Employee training and education, which should consist of: a. Basic training by flow cytometer manufacturers and additional training in hands-on workshops for flow cytometer operators and supervisors. b. Education of laboratory directors in flow cytometric immunophenotyping through workshops and other programs. c. Continuing education in new developments for all flow cytometric immunophenotyping personnel through attendance at meetings and workshops. d. Adherence to federal and state regulations for training and education. 7. Assurance of satisfactory performance. Laboratories must successfully participate in a performance evaluation program. When proficiency testing programs are approved by HCFA as meeting the requirements of CLIA '88 (none are currently approved for CD4+ T-cell testing), laboratories must satisfactorily participate. 8. Review and revision (as necessary, or at established intervals) of the laboratory's policies and procedures to assure adherence to the quality assurance program. All staff involved in the testing should be informed of any problems identified during the quality assurance review, and the corrective actions should be taken to prevent recurrences. B. Document all quality assurance activities.
*Suggested three-color panels. (See Section VI.C.)
Table of Contents This article was provided by U.S. Centers for Disease Control and Prevention. It is a part of the publication Morbidity and Mortality Weekly Report. |
|